Protocol
actions

Find protocols from experts, make them your own


UCLA Health

Los Angeles, United States
Elements Spine SRS

Spine SRS 4.0 Protocol 30Gy 5x

This protocol is originally from Elements Spine SRS 1.5 and it is another simplistic protocol. Devel…
This protocol is originally from Elements Spine SRS 1.5 and it is another simplistic protocol. Developed for 30 Gy in 5 fractions. The maximum dose limit is 36 Gy to 0.1% volume. Most important OAR is the spinal cord. Desired coverage is 95%. Controlled inhomogeneity is not turned on. Dose grid resolution 1mm, Monte Carlo (MC) special resolution 2mm, optimization MC uncertainty is 4% and final MC forward calculation uncertainty is 2%.
Disclaimer:
Please note that this treatment planning template is only a suggestion made by other Novalis Circle Users and is not a medical advice nor a substitute for professional, individual medical diagnosis and/or judgement. The use of the protocol and validation of data are at your own risk and responsibility. You are solely responsible for any treatment decisions that are based on the information contained in the protocol. You acknowledge and agree that the protocol may contain errors and/or omissions and is provided on a strictly “as is” basis without warranties or representations of any kind, whether express or implied, including, but not limited to, the implied warranties of merchantability, accuracy of data, non-infringement or fitness for a particular purpose. To the extent permitted by applicable law, neither Brainlab nor the User that provided the protocol assumes no liability whatsoever resulting from the use of the protocol.